|本期目录/Table of Contents|

[1]魏凌轩,葛斌△,张磊,等.一种新型机械灌注泵的性能模拟[J].生物医学工程研究,2019,01:120-123.
 WEI Lingxuan,GE Bin,ZHANG Lei,et al.Performance simulation on a new type of machine perfusion pump[J].Journal of Biomedical Engineering Research,2019,01:120-123.
点击复制

一种新型机械灌注泵的性能模拟(PDF)

《生物医学工程研究》[ISSN:1006-6977/CN:61-1281/TN]

期数:
2019年01期
页码:
120-123
栏目:
出版日期:
2019-03-25

文章信息/Info

Title:
Performance simulation on a new type of machine perfusion pump
文章编号:
1672-6278 (2019)01-0120-04
作者:
魏凌轩1葛斌1△张磊1张少伟1伍进平1方旭晨2△
1.上海理工大学 医疗器械与食品学院,上海 200093;2.上海市杨浦区市东医院,上海200092
Author(s):
WEI Lingxuan1 GE Bin1 ZHANG Lei1 ZHANG Shaowei1 WU Jinping1 FANG Xuchen2
1.School of Medical Instrument and Food Engineering,University of Shanghai for Science and Technology,Shanghai 200093,China;2.Shidong Hospital,Shanghai 200092
关键词:
机械灌注CFD模拟瞬态分析运动方式质量流量
Keywords:
Machine perfusion Pump CFD simulation TransientAnalysisOperation modeMass flow rate
分类号:
R318
DOI:
10.19529/j.cnki.1672-6278.2019.01.25
文献标识码:
A
摘要:
介绍一种新型机械灌注泵的结构与工作原理,通过CFD模拟对该泵的工作过程进行了仿真。结果显示:液体流入泵时会发生扩散,液体速度从对称轴处往外逐渐减小;液体流出泵时速度分布均匀;液体流动的湍流动能很小;泵的出口质量流量和永磁体端面受力大小由永磁体的运动方式决定,当永磁体运动的加速度以分段函数变化时,出口质量流量能长时间保持稳定;当加速度以正弦函数变化时,流出液体的搏动性更显著,并且流量大小能被更准确得控制。
Abstract:
To introduce the structure and working principle of the new machine perfusion pump. CFD method is utilized to analyze the working process of the new pump. The result shows that the liquid diffuses when the liquid flows into the pump, and the velocity of the liquid gradually deceases outward from the axis of symmetry, the liquid has a uniform velocity distribution when the liquid flows out of the pump, and the turbulent kinetic energy of liquid flow is very small. The rate of the outlet mass flow and the force on the end face of permanent magnet is determined by the motion of the permanent magnets. When the acceleration of the motion of the permanent magnet changes as a piecewise function, the mass flow rate of the pump can remain stable for a long time. When the acceleration of the motion of the permanent magnet changes as a sine function, the pulsation of the liquid is more pronounced, and the mass flow rate can be more accurately controlled.

参考文献/References

备注/Memo

备注/Memo:
(收稿日期:2018-05-01)△通信作者Email:gb13992@hotmail.com;enjoyv6715@hotmail.com
更新日期/Last Update: 2019-05-26