[1]李文慧,刘子悦,杨玺霖,等.用于疤痕治疗的能量可控等离子设备[J].生物医学工程研究,2022,02:208-215.
LI Wenhui,LIU Ziyue,YANG Xilin,et al.An energy controllable plasma equipment for scar treatment[J].Journal of Biomedical Engineering Research,2022,02:208-215.
点击复制
《生物医学工程研究》[ISSN:1006-6977/CN:61-1281/TN]
- 期数:
-
2022年02期
- 页码:
-
208-215
- 栏目:
-
- 出版日期:
-
2022-06-25
文章信息/Info
- Title:
-
An energy controllable plasma equipment for scar treatment
- 文章编号:
-
1672-6278 (2022)02-0208-08
- 作者:
-
李文慧1; 刘子悦1; 杨玺霖1; 湛梦莹1; 侯振中1; 蒋松怿1; 赵博1; 刘玉明1; 游国鹏2; 周宇1△
-
1.上海理工大学医疗器械与食品学院,上海 200093;2.西藏自治区第二人民医院,拉萨 850030
- Author(s):
-
LI Wenhui1; LIU Ziyue1; YANG Xilin1; ZHAN Mengying1; HOU Zhenzhong1; JIANG Songyi1; ZHAO Bo1; LIU Yuming1; YOU Guopeng2; ZHOU Yu1
-
1.College of Medical Instrument and Food Engineering,University of Shanghai for Science and Technology,Shanghai 200093,China;2.Second People’s Hospital of Tibet Autonomous Region,Lhasa 850030,China
-
- 关键词:
-
疤痕; 精细化治疗; 能量控制; 等离子; 电离; 马铃薯
- Keywords:
-
Scar; Refined treatment; Energy control; Plasma; Ionization; Potato
- 分类号:
-
R318;TH772;TH777
- DOI:
-
10.19529/j.cnki.1672-6278.2022.02.16
- 文献标识码:
-
A
- 摘要:
-
本研究设计了一款基于能量控制的便携式等离子设备用于疤痕治疗。为验证该装置基于能量控制的电离效果,以马铃薯为实验材料,设置0.005、0.03、0.1 J的电离实验,观察电离后马铃薯块,分析等离子能量的作用效果。结果显示,0.005 J的电离率为90%;0.03、0.1 J的电离率为100%。随着能量增加,正面和剖面的氧化面积持续增大,配对样本t检验结果显示,0.005、0.03、0.1 J的氧化面积两两之间均存在显著性差异(P<0.05);剖面氧化深度和宽度增大,但氧化宽度随能量增大到0.03 J,逐渐趋于稳定;氧化深度和宽度的比值随能量上升而增大。说明等离子能量的作用深度随能量增加持续变深。
- Abstract:
-
We designed a portable plasma device based on energy control for scar treatment. In order to verify the ionization effect of the device based on energy control, potato was used as the experimental material, and ionization experiments of 0.005,0.03 and 0.1 J were set up to observe the potato pieces after ionization and analyze the effect of plasma energy.The results showed that the ionization rate of 0.005 J was 90%, and the ionization rate of 0.03 and 0.1 J was 100%. The oxidation area of the front and cross section continued to increase with the increase of energy.The paired sample t test results showed that there was a significant difference between the oxidation area of 0.005, 0.03 and 0.1 J (P<0.05); the oxidation depth and width of the profile increased,but the oxidation width was gradually stable as the energy increased to 0.03 J; The ratio of oxidation depth to width increased with the increase of energy. It indicates that the depth of action of plasma energy continues to deepen as the energy increases.
备注/Memo
- 备注/Memo:
-
(收稿日期:2021-11-04)△通信作者Email:zhouyu_working@163.com